Login | Contact | Distributor contact | Imprint
  • Home
  • Search
  • News
  • Downloads
  • Catalogue
  • Register
SHOP ONLINE



  • Applications
  • Products
Proteins, Peptides and Amino acids
Nucleic Acids
Saccharides and Organic Acids
Drugs, Metabolites and Chiral Compounds
Vitamins, Hormones and Lipids
Pesticide Residue, Endocrine Disruptors and Environmental Poll
Synthetic Polymers
Anions and Cations
Column Switching
    • Ratio of Deproteinization (PK-2A)
    • Recovery of Various Drugs (PK-2A)
    • Analysis of Caffeine in BSA
    • Examination of Pretreatment Conditions (1) Dependency on Flow Rate
    • Examination of Pretreatment Conditions (2) Dependency on Quantity of Eluent
    • On-line surfactant removal (MSpak PK&GF)
    • BSA analysis in surfactant (on-line surfactant removal)
    Other Applications
    NP & RP (Normal & Reversed Phase)
    SEC (Size Exclusion Chromatography)
    Saccharides & Organic Acids Analysis
    IEX (Ion Exchange Chromatography)
    IC (Ion Chromatography)
    HIC (Hydrophobic Interaction Chromatography)
    Affinity Chromatography
    Chiral Separation
    LC/MS & Rapid Analyses of Medicines
    Exclusive Use
    Calibration Standards for SEC
    Column Switching
    GPC Clean-up
    Detectors
    New products
    Download

    Recovery of Various Drugs (PK-2A)

    Recoveries of the various drugs by MSpak PK-2A ( a Column for Column switching ) were investigated. PK series can concentrate various drugs over the wide range of pH and hydrophobicity.



    Sample : Drugs 2.5ppm each, 100µL

    Pretreatment
    Column         : Shodex MSpak PK-2A (2.0mmID*10mm)
    Eluent         : 10mM CH3COONH4(pH7.0)
    Flow rate      : 0.5mL/min
    Column temp.   : Room temp.
    Switching time : 5min
     
    Analysis 
    Column       : Shodex MSpak GF-4A (4.6mmID*10mm)
    Eluent       : 10mM CH3COONH4(pH7.0)/CH3CN
    Flow rate    : 0.5mL/min
    Column temp. : Room temp.

     
    © Alle Rechte liegen bei Shodex®