Effects of Acetonitrile Concentration on Elution Time (NH2P-50)

The separation of saccharides using Asahipak NH2P-50 column can be controlled by changing mixing ratio of acetonitrile and water in the eluent. Since the NH2P-50 column is free from non-specific adsorption, sharp and symmetrical peaks can be obtained for most of saccharides.
When the NH2P-50 column is used, as same as when silica-based amino columns are used, the retention time of saccharides becomes longer as the acetonitrile concentration in the eluent becomes higher. The elution order of saccharides using the NH2P-50 columns is almost as same as that using silica-based amino columns though there are some exception such as the order between galactose and glucose and between lactose and maltose.

Sample :
1. Rhamnose
2. Xylose
3. Fructose
4. Glucose
5. Sucrose
6. Maltose
7. Lactose
8. Galactose
9. Raffinose
10. Mannose



Columns : Shodex Asahipak NH2P-50 4E (4.6mmI.D. x 250mm) , Silica based amino
Column(other manufacturer)
Eluent : CH3CN/H2O
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 30°C