+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Effects of Sample Solvent Composition

Effects of sample solvent composition when Asahipak NH2P-50 is used is shown. When making sample solution to be injected, sample should be dissolved in a solvent whose composition is as close as the eluent, preferably in the eluent. When 75% acetonitrile solvent is used for the eluent and a sample cannot be dissolved in it, first, the sample should be dissolved in a solvent which dissolves the sample, and then, adjust the solvent composition to be acetonitrile aqueous solution of 50% or more.

Sample : 1mg/mL each, 20μL
1. meso-Erythritol 2. Xylitol 3. Fructose 4. Glucose 5. Sucrose 6. Lactose 7. Maltose



Column : Shodex Asahipak NH2P-50 4E (4.6mmI.D. x 250mm)
Eluent : CH3CN/H2O=75/25
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 25°C