+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Monosaccharides (2) (NH2P-50 4E)

Monosaccharides were analyzed using Asahipak NH2P-50 4E, a
Column for saccharides analysis. Fucose and rhamnose can be separated by adding phosphoric acid to the
Eluent.

Sample : 0.5% each, 10μL
1. Rhamnose
2. Fucose
3. Xylose
4. Arabinose
5. Mannose
6. Glucose
7. Galactose

Column : Shodex Asahipak NH2P-50 4E (4.6mmI.D. x 250mm)

Eluent : (above); 250mM H3PO4 aq./CH3CN=20/80 (below); H2O/CH3CN=20/80

Flow rate : 1.0mL/min

Detector : Shodex RI

Column temp. : (above); 50°C, (below); 30°C
Equilibration of NH2P-50 4E
Column : After passing 60mM Phosphoric acid in the
Column by 0.5 mL/min for 2 hours, it replaces by
Eluent.