+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

LC/MS Analysis of N-Acetylchitooligosaccharides

Sugar analysis
Column, Asahipak NH2P-40 2D, was used with a LC/MS to separate N-Acetylchitooligosaccharides. By adding ammonia to the solvent, it enhances the ionization of saccharides, and thus provides high sensitivity detection.

Sample : 5μL, Mixture of N-Acetylchitooligosaccharides
(Seikagaku Co. code:800104 Chitooligosaccharides Mixture) 500ng/mL
1. N-Acetyl-D-glucosamine
2. di-N-Acetylchitobiose
3. tri-N-Acetylchitotriose
4. tetra-N-Acetylchitotetraose
5. penta-N-Acetylchitopentaose
6. hexa-N-Acetylchitohexaose

Column : Shodex Asahipak NH2P-40 2D (2.0mmI.D. x 150mm)

Eluent : (A); 0.05% Ammonia aq. (B); CH3CN Linear gradient; (B%) 75% (0min) ↁE50% (10min) ↁE50% (15min)

Flow rate : 0.2mL/min

Detector : ESI-MS(SIM Negative)

Column temp. : 30°C