+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Analysis of Cyclodextrins (NH2P-50 4E)

Three cyclodextrins were separated using Asahipak NH2P-50 4E, a polymer-based amino column. These conditions lead to a short elution time with greater resolution compared to the separation using an SEC column.

Sample : 20μL
250μg/mL each (in H2O/CH3CN=50/50)
1. α-Cyclodextrin
2. γ-Cyclodextrin
3. β-Cyclodextrin



Column : Shodex Asahipak NH2P-50 4E (4.6mmI.D. x 250mm)
Eluent : CH3CN/H2O=60/40
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 40°C