Feature of High Temperature Analysis (3) (Fast Analysis)

ET-RP1 4D is polymer base column for high temperature analysis. Figure shows comparison chromatograms between 40°C and 80°C. Mobile phase viscosity decreased at high temperature analysis, higher flow rate can be applied. And suitable flow rate range is shifted to higher at high temperature. ET-RP1 4D is good column for fast analysis with high temperature analysis.
Please refer to


Sample : 5μL, 10mg/L each
1. Caffeine
2. Pyridine
3. Phenol
4. Propyl paraben
5. Acetophenone
6. Benzene
7. Toluene
(notes)
* Eluent was introduced to column after preheating and introduced to detector after cooling.
* Preheat eluent when high flow rate (more than 1mL/min) is applied.

Column : Shodex ET-RP1 4D (4.6mmI.D. x 150mm)
Eluent : CH3CN/H2O=40/60
Detector : UV(210nm) *
Column oven : Polaratherm 9000 Series (SandraSelerity Technologies, Inc)