+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Analysis of Voglibose According to JP Method (NH2P-50 4E)

According to JP (Japanese Pharmacopoeia) method, purity test of voglibose, it is necessary for the system suitability to satisfy theoretical plate number of ≥ 7,000 and symmetry factor of 0.8 -1.2, and for the system repeatability to satisfy relative standard deviation (RSD) of ≤ 3.0% (repeated six times). It was confirmed that the all conditions were satisfied when voglibosel was analyzed using Asahipak NH2P-50 4E.

Sample : 10μg/mL, 50μL
1. Voglibose

Column : Shodex Asahipak NH2P-50 4E (4.6mmI.D. x 250mm)

Eluent : 20mM Sodium phosphate buffer(pH6.5)/CH3CN=37/63 Reagent : 12mM NaIO4 + 50mM Taurine aq.

Flow rate : (

Eluent) 0.6mL/min (Reagent) 0.5mL/min

Detector : Fluoresence(Ex. 350nm, Em. 430nm) (post-

Column reaction)

Column temp. : 25°C Reaction temp. : 100°C Cooling temp. : 15°C
*For
Column equilibration: First introduce 100mM Sodium phosphate buffer(pH6.5) at 0.5 mL/min for 2 hours, and then switch to the
Eluent at 0.5 mL/min for 2 hours.