+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Human Serum (4) (DEAE-825)

Print
Human serum was analyzed using IEC DEAE-825 (a column for waek anion exchange chromatography).


Sample : Human serum

Column       : Shodex IEC DEAE-825 (8.0mmI.D. x 75mm)
Eluent       : (A); 50mM Tris-HCl buffer(pH8.0)
               (B); (A) + 0.5M NaCl 
               Linear gradient: 0min to 60min, (A) to (B)
Flow rate    : 1.0mL/min
Detector     : UV(280nm)
Column temp. : Room temp.