+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

LC/MS Analysis of Metformin (NH2P-40 2D)

High polar compounds, such as metformin (an oral antidiabetic drug in biguanide class) are not retained by reversed-phase chromatography. Instead, HILIC mode is suitable for the analysis of polar compounds. A polymer-based amino column, Asahipak NH2P-40 2D, was used in the LC/MS method. Eluent containing a high acetonitrile ratio provides a high sensitive MS detection.

Sample : 5μL
1. Metformin -HCl 5ng/mL
(in H2O/CH3CN=5/95)



Column : Shodex Asahipak NH2P-40 2D (2.0mmI.D. x 150mm)
Eluent : 50mM HCOONH4 aq./CH3CN=5/95
Flow rate : 0.2mL/min
Detector : ESI-MS (SIM)
Column temp. : 30°C