+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Rain Water (NI-424)

Print
Four kinds of anions in rain water were analyzed using IC NI-424 (an anion analysis column for non-suppressor method).


Sample : Rain water, 100μL
1. Cl-
2. NO2-
3. NO3-
4. SO42-
Column       : Shodex IC NI-424 (4.6mmI.D. x 100mm)
Eluent       : 8mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid 
               + 0.005mM *CyDTA aq. 
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40°C
* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid