+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Anion Analysis in Lithium Salts (1) (NI-424)

Print

Lithium tetrafluoroborate, lithium hexafluorophosphate and lithium perchlorate are used as an electrolyte of a lithium ion secondary battery. In this application, the anions of these salts were analyzed using IC NI-424, an anion analysis column for non-suppressor method.


Sample : 20μL
1. BF4- 20mg/L
2. ClO4- 50mg/L
3. PF6- 50mg/L


Column       : Shodex IC NI-424 (4.6mmI.D. x 100mm)
Eluent       : 8mM 4-Hydroxybenzoic acid + 3.4mM Bis-Tris + 2mM Phenylboronic acid
               + 0.005mM *CyDTA aq.
Flow rate    : 1.0mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40℃

* CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid