+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, München Mo-Do 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, München   

Melamine and Related Substances (VG-50 4D)

Melamine, ammeline, ammelide, and cyanuric acid were analyzed by HILIC mode simultaneously using polymer-based amino
Column HILICpak VG-50 4D.

Sample : 5μL
5μg/mL each
(in H2O/CH3CN=50/50)
1. Melamine
2. Ammeline
3. Cyanuric acid
4. Ammelide    

Column : Shodex HILICpak VG-50 4D (4.6mmI.D. x 150mm)

Eluent : 30mM Ammonium formate/CH3CN=35/65

Flow rate : 0.6mL/min

Detector : Corona charged aerosol

Column temp. : 40°C