Polyacrylamide (2) (KS-807, SB-806 HQ)

The analytical results of polyacrylamide are shown. In the analysis with OHpak SB-806 HQ (exclusion limit: 2.0x107g/mol), polyacrylamide elutes around exclusion limit. On the other hand, in the analysis with SUGAR KS-807 (exclusion limit: 2.0x108g/mol), polyacrylamide elutes with SEC mode.

Sample : 100μL
0.02% Polyacrylamide (PAAm)
(Scientific Polymer Product)


Columns : Shodex SUGAR KS-807, OHpak SB-806 HQ (8.0mmI.D. x 300mm each)
Eluent : 0.2M NaCl aq.
Flow rate : 0.5mL/min
Detector : Shodex RI and MALS(Multi angle laser light scattering
Detector)
Column temp. : 30°C