+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, Munich Mo-Thu 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, Munich   

Comparison of SEC Separation of IgG between LW-403 4D and LW-803

Print

PROTEIN LW-403 4D is an aqueous SEC (GFC) semi-micro column for rapid analysis of antibody proteins. PROTEIN LW-403 4D and PROTEIN LW-803 (a standard column) were compared for the chromatogram of IgG. The analysis time using LW-403 4D can be completed in half the time compared to LW-803.
 

Sample : IgG from human serum 10mg/mL
(LW-403 4D) 0.5μL
(LW-803) 5μL

(D) Dimer
(M) Monomer


Columns      : Shodex PROTEIN LW-403 4D (4.6mmI.D. x 150mm)
               Shodex PROTEIN LW-803 (8.0mmI.D. x 300mm)
Eluent       : 50mM Sodium phosphate buffer (pH7.0) + 0.3M NaCl
Flow rate    : (LW-403 4D) 0.35mL/min
               (LW-803)    1.0mL/min
Detector     : (LW-403 4D) UV(280nm) (small cell volume)
               (LW-803)    UV(280nm) (conventional type)
Column temp. : Room temp.