+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, Munich Mo-Thu 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, Munich   

Analysis of Tagatose According to USP-NF Method (SC1011)

Print

According to USP40-NF35 method, the assay of tagatose should be used a column packed with L22. It is necessary for the system suitability to satisfy relative standard deviation (RSD) of ≤ 2.0%. It was confirmed to satisfy the system suitability when tagatose was analyzed using SUGAR SC1011.


Sample : 20 µL
1. Tagatose 5 mg/mL


Column       : Shodex SUGAR SC1011 (8.0 mm I.D. x 300 mm)
Eluent       : 0.05 mg/mL Calcium acetate aq.
Flow rate    : 0.6 mL/min
Detector     : Shodex RI
Column temp. : 85 °C