+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, Munich Mo-Thu 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, Munich   

Analysis of Choline and Acetylcholine (DE-413, ACH-494)

Print
Choline and acetylcholine separated by RSpak DE-613 or DE-413 are converted to hydrogen peroxide (H2O2) by AFpak ACH-494, a column with choline oxidase and acetylcholine esterase immobilized. This hydrogen peroxide can be detected by an electrochemical detector (ECD). We recommend that DE-413 is used to reduce analytical time.



Sample :
1. CHO (Choline )
2. EHC (Ethylhomocholine )
3. ACH (Acetylcholine )
Columns      : Shodex RSpak DE-613 (6.0mmI.D. x 150mm), DE-413 (4.6mmI.D. x 150mm)
Post column  : Shodex AFpak ACH-494 (4.6mmI.D. x 10mm)
Eluent       : 0.1M Phosphoric acid + 300mg/L Sodium 1-decansulfonate
               + 65mg/L Tetramethylammonium chloride(pH8.0 adjusted by 1M NaOH)
Flow rate    : 1.0mL/min
Detector     : ECD (Electrode: Pt, 350mV SCE)
Column temp. : 37°C