+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, Munich Mo-Thu 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, Munich   

Chiral Separation of Mandelic Acid (CRX-853)

Print

A column for chiral separation, ORpak CRX-853, was used to separate D and L isomers of mandelic acid.


Sample : DL-Mandelic acid 0.01%, 20μL
1. D-Mandelic acid
2. L-Mandelic acid


Column       : Shodex ORpak CRX-853 (8.0mmI.D. x 50mm)
Eluent       : 0.25mM CuSO4 aq.
Flow rate    : 1.0mL/min
Detector     : UV(230nm)
Column temp. : 50°C