Styrene butyl methacrylate copolymer was analyzed using GPC LF-404 (a column for organic SEC (GPC)).
Please refer to OUTLINE of LF-804.


Sample : 0.2% Styrene butyl methacrylate copolymer, 10μL



Columns      : Shodex GPC LF-404 (4.6mmI.D. x 250mm) x 2
Eluent       : THF
Flow rate    : 0.3mL/min
Detector     : Shodex RI(small cell volume)
Column temp. : 40°C