+49 89 9399 6237 info(at)shodex.de Konrad-Zuse-Platz 3, Munich Mo-Thu 8:30 - 17:00 / Fr 8:30 - 16:00
Konrad-Zuse-Platz 3, Munich   

Analysis of Hydrophobic Anions (I-524A)

Print

Four hydrophobic anions were analyzed using an anion analysis column for non-suppressor method, IC I-524A. The method optimized here is able to analyze hydrophobic anions without the interference of standard anions.


Sample : 20μL
1. Cl- 5mg/L
2. NO3- 10mg/L
3. SO42- 10mg/L
4. S2O32- 50mg/L
5. I- 50mg/L
6. SCN- 50mg/L
7. ClO4- 50mg/L


 

Column       : Shodex IC I-524A (4.6mmI.D. x 100mm)
Eluent       : 2.5mM Phthalic acid + 3.3mM Tris(hydroxymethyl)aminomethane aq.
Flow rate    : 1.2mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40°C