Bromate in Drinking Water (SI-50 4E)
Please use this column with suppressor type ion chromatography system. Recently ozonation has been applied for sterilization of drinking water. Ozonation transforms part of bromide to bromate which is a potential carcinogen. So, it is very important to measure bromate in drinking water. A modeldrinking water with bromate was measured by suppressed coductivity with IC SI-50 4E.

Sample : Model drinking water, 500μL
1. 0.1mg/L F
- 2. 0.003mg/L BrO
3- 3. 20mg/L Cl
- 4. 0.1mg/L Br
- 5. 10mg/L NO
3- 6. 10mg/L SO
42-
1. : 0.02 mg/L
2. : 0.01 mg/L
3. : 0.003mg/L
Column : Shodex IC SI-50 4E (4.0mmI.D. x 250mm)
Eluent : 3.2mM Na2CO3 + 1mM NaHCO3 aq.
Flow rate : 0.8mL/min
Detector : Suppressed conductivity
Column temp. : 25°C