The polymer-based SEC (GFC) columns with additional separation mode
Product information
Multi mode chromatography is an analytical method combining several separation modes.
When a single separation mode is not enough for separating the target substance, an improvement can be expected by using multi mode chromatography. The balance between the separation modes involved can be controlled by changing the eluent composition.
Asahipak GS series [multi-mode columns]
-
Polymer-based columns
-
SEC is the main separation mode
-
With the choice of eluent, the column provides multimode features of reversed phase, HILIC, and ion exchange modes to SEC
-
Suitable for the separation of peptides or nucleic acids with similar molecular weights
-
Suitable for desalting samples or substituting buffer in protein analysis
The base material is a polymer, resulting in the following advantages;
1) usable at a wide pH range, 2) washable with alkaline solutions, 3) column durability.
Asahipak GS columns operate basically in SEC mode. However, depending on the eluent conditions, it allows separation by multiple modes combining reversed phase and ion exchange modes. The column is particularly effective in analyzing samples such as proteins and nucleic acid components, which are made up of both acidic and basic components. In this separation mode, samples are eluted in the order of acidic, neutral and basic. Separation can also be adjusted by changing the eluent components.
1) When a neutral/acidic eluent is used.
Acidic samples: As it is the ion exclusion mode which is in operation, substances are eluted relatively quickly.
Neutral samples: As substances are separated by reversed phase mode, they are eluted moderately.
Basic samples: As it is the ion exchange mode which is in operation, substances are eluted relatively slow.
2) When a basic eluent is used.
Acidic samples: Basically, substances are excluded together at the V(0) position but, with acidic samples of large molecular weights, there are cases in which substances are retained in the column.
Neutral samples: Substances are separated by reversed phase mode.
Basic samples: If the eluent has a higher pH than the pKa of the sample, substances are separated by reversed phase mode as the dissociation of basic samples is suppressed.
Product overview
Analytical Columns
Product code | Product name | Separation | Functional group | Gel material | Plates per column | Particle size | Pore size | Size (ID x length) | Housing material |
---|---|---|---|---|---|---|---|---|---|
F6710019 | Asahipak GS-2G 7B | SEC | - | Polyvinyl alcohol | (guard) | 9 µm | - | 7.5 x 50 mm | steel |
F7600006 | Asahipak GS-320 HQ | SEC | - | Polyvinyl alcohol | ≥ 19,000 | 6 µm | 400 Å | 7.5 x 300 mm | steel |
F7600005 | Asahipak GS-220 HQ | SEC | - | Polyvinyl alcohol | ≥ 19,000 | 6 µm | 150 Å | 7.5 x 300 mm | steel |
Preparative Columns
Product code | Product name | Separation | Functional group | Gel material | Plates per column | Particle size | Pore size | Size (ID x length) | Housing material |
---|---|---|---|---|---|---|---|---|---|
F6810034 | Asahipak GS-220 20G | SEC | - | Polyvinyl alcohol | ≥ 14,000 | 13 µm | 150 Å | 20.0 x 500 mm | steel |
F6810035 | Asahipak GS-320 20G | SEC | - | Polyvinyl alcohol | ≥ 14,000 | 13 µm | 400 Å | 20.0 x 500 mm | steel |
Product details
Analytical Columns
Product name | Maximum pressure | Usual flow rate | Maximum flow rate | Temperature range | pH range | Salt concentration | Shipping solvent | USP | MW range | Exclusion limit |
---|---|---|---|---|---|---|---|---|---|---|
Asahipak GS-2G 7B | - | - | - | 4 to 60°C | 2 to 12 | ≤ 0.5 M | H2O/CH3OH = 70/30 | - | - | - |
Asahipak GS-320 HQ | 5.0 MPa (50 bar) | 0.4 to 0.6 mL/min | 1.0 mL/min | 4 to 60°C | 2 to 12 | ≤ 0.5 M | H2O/CH3OH = 70/30 | - | 300 - 20,000 | 40,000 |
Asahipak GS-220 HQ | 6.0 MPa (60 bar) | 0.4 to 0.6 mL/min | 1.0 mL/min | 4 to 60°C | 2 to 9 | ≤ 0.5 M | H2O/CH3OH = 70/30 | - | 300 - 3,000 | 7,000 |
Preparative Columns
Product name | Maximum pressure | Usual flow rate | Maximum flow rate | Temperature range | pH range | Salt concentration | Shipping solvent | USP | MW range | Exclusion limit |
---|---|---|---|---|---|---|---|---|---|---|
Asahipak GS-220 20G | - | 4.0 to 6.0 mL/min | 8.0 mL/min | 4 to 60°C | 2 to 9 | ≤ 0.5 M | H2O/CH3OH = 70/30 | - | 300 - 3,000 | 7,000 |
Asahipak GS-320 20G | - | 5.0 to 8.0 mL/min | 12 mL/min | 4 to 60°C | 2 to 12 | ≤ 0.5 M | H2O/CH3OH = 70/30 | - | 300 - 20,000 | 40,000 |
Product info
Applications
At the moment a new window opens and you are transferred to our Japanese website, when you click on the application. We are working on a better solution.
Calibration Curves for GS-HQ Series (Aqueous Eluent: PEG, PEO)
Calibration Curves for GS-HQ Series (Aqueous Eluent: Pullulan)
Eluent Conditions for Asahipak GF, GS HQ series
Low-Molecular-Weight Water Soluble Dietary Fiber (GS-220 HQ, KS-802)
Poly(Ethylene Glycol) Standards (9) (GS-220 HQ)
Polyphosphates (3) (GS-220 HQ)
Polyphosphates (4) (GS-220 HQ)
Pullulan Standards (7) (GS-220 HQ)
Adenine Nucleotides in Red Blood Cell
Analysis of ATP and its Related Substances in Muscle of Fish and Shellfish (GS-320 HQ)
Analysis of Peptide using Multimode Column
Chymotrypsin-Hydrolysated Peptides of Ascidian 29K Protein
Effects of Analysis of PEG Mixtures with GS-320 HQ by Composition of Eluent
Hydroxyriboflavin and Riboflavin in Urine
Influence of Column Temperature on Retention Time of PEG (GF-310 HQ and GS-320 HQ)
Isolation of Ecdyson from Vitex Fiherii
Nucleobases, Nucleosides and Nucleotides (GS-320 HQ)
Peptides in Extract from Waste Water Sludge
Preparation of CS-Glucan in Coix Seed
Purine Base in Beer (GS-320 HQ)
Purine Bases and Uric Acid Separation (GS-320 HQ)
Recovery of Proteins (GF-310 HQ and GS-320 HQ)
Related products
Shodex STANDARD P-82 (Pullulan) is a calibration standard kit for aqueous size exclusion chromatography.
Pullulan is an unbranched polysaccharide: units of maltotriose are connected with α-1,6 bond.
-
Easy to prepare
-
For aqueous SEC (GFC)
-
Unbranched pullulan standard
-
High solubility in water eliminates the possibility of recrystallization
STANDARD P-82 kit consists of 8 vials with different molecular weight standards of 200 mg each:
MW 800.000, 400.000, 200.000, 100.000, 50.000, 20.000, 10.000, 5.000