top of page

Column cleaning and troubleshooting


When a column is deteriorated by the foreign substances remaining in the column or adsorbed by the packing material, the substances might be washed out by the procedure described below. However, sometimes the procedure is not sufficient to regenerate the column and, in such case, it is necessary to replace the column with a new one. And, even if the column can be regenerated, the column performance may be not so good as before.

 

After cleaning the column, be sure to replace the in-column solvent to the normal solvent. The use of a guard column at the up-stream of the main column is recommended to prevent the main column from the deterioration by the foreign substances. When a guard column is used, first, the guard column will be deteriorated by the foreign substances and it is possible to prevent the main column from the deterioration by replacing the guard column regularly.


When cleaning a guard column, be sure to disconnect the main column before the cleaning. If a guard column and a main column are connected in series and the cleaning procedure is performed, foreign substances in the guard column may move to the main column and, as the consequence, the main column may be deteriorated.

For more information, please also read and download the column manuals:

Cleaning procedure and regeneration of  Reversed Phase (RP) columns

Asahipak ODP-50 series

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.

  • Method 1: Basic solvent wash

Apply after using the column under acidic to neutral condition for an extended time period. Introduce 0.1 M aqueous sodium hydroxide solution/acetonitrile = 80/20.

  • Method 2: Acidic solvent wash

Apply after using the column under basic to neutral condition for an extended time period. Introduce 50 mM aqueous phosphoric acid solution/acetonitrile = 90/10.

  • Method 3: Organic solvent wash

Apply after using the column under highly aqueous condition for an extended time period. Introduce water/acetonitrile = 20/80.

Asahipak C4P

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.

  • Method 1: Basic solvent wash

Apply after using the column under acidic to neutral condition for an extended time period. Introduce 0.1 M aqueous sodium hydroxide solution/acetonitrile = 80/20.

  • Method 2: Acidic solvent wash

Apply after using the column under basic to neutral condition for an extended time period. Introduce 50 mM aqueous phosphoric acid solution/acetonitrile = 90/10.

  • Method 3: Organic solvent wash

Apply after using the column under highly aqueous condition for an extended time period. Introduce water/acetonitrile = 20/80.

RSpak DE series

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.

  • Method 1: Basic solvent wash

Apply after using the column under acidic to neutral condition for an extended time period. Introduce 10 mM aqueous sodium hydroxide solution/acetonitrile = 80/20.

  • Method 2: Acidic solvent wash

Apply after using the column under basic to neutral condition for an extended time period. Introduce 50 mM aqueous phosphoric acid solution/acetonitrile = 90/10.

  • Method 3: Organic solvent wash

Apply after using the column under highly aqueous condition for an extended time period. Introduce water/acetonitrile = 20/80.

RSpak DS series

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.

  • Method 1: Basic solvent wash

Apply after using the column under acidic to neutral condition for an extended time period. Introduce 10 mM aqueous sodium hydroxide solution/acetonitrile = 80/20.

  • Method 2: Acidic solvent wash

Apply after using the column under basic to neutral condition for an extended time period. Introduce 50 mM aqueous phosphoric acid solution/acetonitrile = 90/10.

  • Method 3: Organic solvent wash

Apply after using the column under highly aqueous condition for an extended time period. Introduce water/acetonitrile = 20/80.

RSpak DM-614

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction.

Method: Set the flow rate at 0.3 mL/min, and introduce 50 mL of methanol, acetonitrile or water.

RSpak NN-814

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.

(2) In case adsorption of ionic substances is suspected, inject 20-50 μL of 2 M nitric acid aqueous solution several times while running the eluent of analysis.

RSpak JJ-50 2D

Reverse the direction and flow rate is 0.1 mL/min.

  • (Pollution by low valency hydrophilic ions)

1) H2O 6 min, 2) 0.1 M NaOH aq. 75 min, 3) H2O 12 min, 4) Eluent 40 min

  • (Pollution by metal ions)

1) H2O 6 min, 2) 0.1 M HNO3 aq. 75 min, 3) H2O 6 min, 4) 0.1 M NaOH aq. 75 min, 5) H2O 12 min, 6) Eluent 40 min

  • (Pollution by high valency hydrophobic ions)

1) H2O 25 min, 2) 5 % acetonitrile aq. 20 min, 3) 100 % acetonitrile 100 min, 4) H2O 50 min

ODP2 series

See column manual for basic, acidic or organic solvent wash.

Silica C18M series

(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.

  • Apply after using the column under highly aqueous conditions. Introduce water/acetonitrile = 20/80.

  • Apply to samples containing protein. Introduce water/acetonitrile = 50/50 - 30/70 (containing 0.1 % TFA)

Cleaning procedure and regeneration of
HILIC columns

Asahipak NH2P-50 4E

(Standard cleaning)
Reverse the direction and flow rate is 0.5 mL/min.
1) H2O 10 min, 2) 0.1 M NaOH aq. 120 min, 3) H2O 20 min, 4) Eluent 60 min
(Cleaning metal ions)
Reverse the direction and flow rate is 0.5 mL/min.
1) H2O 10min, 2) 0.1 M HClO4 aq. 60 min, 3) H2O 10 min, 4) 0.1 M NaOH aq. 120 min, 5) H2O 20 min, 6) Eluent 60 min

HILICpak VG-50

(Standard cleaning) VG-50 4E
Reverse the direction and flow rate is 0.5 mL/min.
1) H2O 10 min, 2) 0.1 M NaOH aq. 120 min, 3) H2O 20 min, 4) Eluent 60 min
(Cleaning metal ions)
Reverse the direction and flow rate is 0.5 mL/min.
1) H2O 10 min, 2) 0.1 M HNO3 aq. 60 min, 3) 1) H2O 10 min, 4) 0.1 M NaOH aq. 120 min, 5) H2O 20 min, 6) Eluent 60 min

(Standard cleaning) VG-50 2D
Reverse the direction and flow rate is 0.1 mL/min.
1) H2O 6 min, 2) 0.1 M NaOH aq. 75 min, 3) H2O 12 min, 4) Eluent 40 min
(Cleaning metal ions)
Reverse the direction and flow rate is 0.1 mL/min.
1) H2O 6 min, 2) 0.1 M HNO3 aq. 75 min 3) 1) H2O 6 min, 4) 0.1 M NaOH aq. 75 min, 5) H2O 12 min, 6) Eluent 40 min

HILICpak VT-50 2D

(Standard cleaning)
Reverse the direction and flow rate is 0.1 mL/min.
1) H2O 6 min, 2) 0.1 M NaOH aq. 75 min, 3) H2O 12 min, 4) Eluent 40 min
(Cleaning metal ions)
Reverse the direction and flow rate is 0.1 mL/min.
1) H2O 6 min, 2) 0.1 M HNO3 aq. 75 min 3) 1) H2O 6 min, 4) 0.1 M NaOH aq. 75 min, 5) H2O 12 min, 6) Eluent 40 min

HILICpak VC-50 2D

Reverse the direction and flow rate is 0.1 mL/min.
1) H2O 10 min, 2) 25 mM HNO3 aq./CH3CH=70/30 60 min, 3) H2O 10 min, 4) Eluent 60 min

HILICpak VN-50

Reverse the direction and flow rate is 0.3 mL/min (VN-50 4D) or 0.1 mL/min (VN-50 2D).
1) H2O 6 min, 2) 0.1 M HNO3 aq. 75 min, 3) H2O 6 min, 4) 0.1 M NaOH aq. 75 min, 5) H2O 12 min, 6) Eluent 40 min

Cleaning procedure and regeneration of
SUGAR columns (Ligand Exchange)

SUGAR SP0810

Flow of H2O at 70 - 95°C, 0.5 mL/min. Then inject of 50 μL of 0.2 M Pb(NO3)2 aqueous 3 or 4 times.

SUGAR SC1011

SUGAR EP SC1011-7F

USPpak MN-431

Flow 50 mL of 0.1 M Ca(NO3)2 aqueous solution at 50°C, 0.5 mL/min.

SUGAR SC1211

After replacing in-column solvent completely with H2O, flow 30 mL of 0.1 M Ca(NO3)2 aqueous solution at 50°C, 0.5 mL/min. Then, replace the in-column solvent completely with H2O and gradually increase the organic solvent concentration to the concentration to be used.

SUGAR KS-800 series

Flow 0.2 M NaOH aqueous solution at 0.5 mL/min. After flowing the solvent, replace the in-column solvent completely with H2O. (Sometimes, injection of 40 μL of 1 M NaOH aqueous solution is sufficient.)

SUGAR SZ5532

After replacing in-column solvent completely with H2O, flow 30 mL of 0.1 M Zn(NO3)2 aqueous solution at 50°C, 0.5 mL/min. Then, replace the in-column solvent completely with H2O and gradually increase the organic solvent concentration to the concentration to be used.

RSpak DC-613

Flow 50 mL of Methanol, Ethanol, CH3CN, or 50 mM NaOH aqueous solution. Or, inject 20 to 50 μL of 0.5 M NaOH aqueous solution.

Cleaning procedure and regeneration of
Ion Exclusion columns (organic acids)

SUGAR SH1011

SUGAR SH1821

Flow 50 mL of 25 mM H2SO4 aqueous solution at 50°C, 0.5 mL/min.

RSpak KC-811

Flow 50 mL of 25 mM H2SO4 aqueous solution at 50°C, 0.5 mL/min.

Cleaning procedure and regeneration of aqueous SEC columns

PROTEIN KW-800

series

Reverse the direction and flow rate is 0.5 mL/min. The volume of the cleaning solvent required is 5 to 10 times the column volume.
1) In case where protein has been adsorbed :
Use a solvent containing 0.5 M salt or phosphate buffer (about pH3.0).
2) In case where a hydrophobic substance has been adsorbed :
Use a solvent containing 10 to 20 % acetonitrile or methanol.

PROTEIN LW-803

Reverse the direction and flow rate is 0.5 mL/min. The volume of the cleaning solvent required is 5 to 10 times the column volume.
1) In case where protein has been adsorbed :
Use a solvent containing 0.5 M salt or phosphate buffer (about pH3.0).
2) In case where a hydrophobic substance has been adsorbed :
Use a solvent containing 10 to 20 % acetonitrile or methanol.

PROTEIN KW-400 series

Reverse the direction and flow rate is 0.15 mL/min. The volume of the cleaning solvent required is 5 to 10 times the column volume.
1) In case where protein has been adsorbed :
Use a solvent containing 0.5 M salt or phosphate buffer (about pH3.0).
2) In case where a hydrophobic substance has been adsorbed :
Use a solvent containing 10 to 20 % acetonitrile or methanol.

PROTEIN LW-403 4D

Reverse the direction and flow rate is 0.15 mL/min. The volume of the cleaning solvent required is 5 to 10 times the column volume.

1) In case where protein has been adsorbed :Use a solvent containing 0.5 M salt or phosphate buffer (about pH3.0).

2) In case where a hydrophobic substance has been adsorbed :Use a solvent containing 10 to 20 % acetonitrile or methanol.

OHpak SB-800 HQ series

OHpak LB-800 series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

Cleaning procedure and regeneration of
organic SEC columns

GPC KF-800 series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

GPC KF-400HQ

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

GPC KD-800 series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

GPC HK-400 series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

GPC LF series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

CLNpak EV series

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. This is effective for the foreign substances remaining in the column, however not effective for the substances adsorbed to the packing material.

MSpak GF-4A

Method 1: Adsorption of hydrophobic compounds (when using aqueous eluent)

For an efficient cleaning, reverse the flow direction. Introduce the eluent with higher acetonitrile or methanol concentration. Set the flow rate at less than 0.2 mL/min. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Method 2: Adsorption of hydrophobic compounds (when using aqueous eluent)

For an efficient cleaning, reverse the flow direction. Introduce the eluent with higher salt concentration. Set the flow rate at less than 0.2 mL/min. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Keep the polar organic solvent and salt concentrations within the concentrations stated in the Operation Manual "List of Applicable Solvents".

Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.

Cleaning procedure and regeneration of
Ion Exchange columns

IEC QA-825

For an efficient cleaning, reverse the flow direction. 

Set the flow rate at 0.5 mL/min. Introduce the eluent and inject 1 to 2 mL of 0.1 M aqueous sodium hydroxide solution several times.

IEC DEAE-825

For an efficient cleaning, reverse the flow direction. 

Set the flow rate at 0.5 mL/min. Introduce the eluent and inject 1 to 2 mL of 0.1 M aqueous sodium hydroxide solution several times.

Asahipak ES-502N 7C

For an efficient cleaning, reverse the flow direction.

Method 1: Adsorption of ionic compounds

Introduce an eluent with higher salt concentration or 10 mM aqueous sodium hydroxide solution. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Method 2: Adsorption of hydrophobic compoundsIntroduce a mixture of solvent containing 50 mM eluent (buffer or aqueous salt solution) and 50 % (v/v) acetonitrile or methanol. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.

IEC SP-825

For an efficient cleaning, reverse the flow direction.

Set the flow rate at 0.5 mL/min. Introduce the eluent and inject 1 to 2 mL of 30 % (v/v) aqueous acetic acid solution several times.

IEC SP-FT 4A

For an efficient cleaning, reverse the flow direction.

Method 1: Adsorption of proteins

Introduce an eluent with higher salt concentration (about 1 M) or 0.1 % nonionic surfactants. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Method 2: Adsorption of hydrophobic compounds

Introduce a mixture of solvent containing 50 mM eluent (buffer or aqueous salt solution) and 10 to 20 % (v/v) acetonitrile. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.

IEC CM-825

For an efficient cleaning, reverse the flow direction.​

Set the flow rate at 0.5 mL/min. Introduce the eluent and inject 1 to 2 mL of 30 % (v/v) aqueous acetic acid solution several times.

Asahipak ES-502C 7C

For an efficient cleaning, reverse the flow direction.

Method 1: Adsorption of ionic compounds

Introduce an eluent with higher salt concentration or 0.1 M aqueous acetic acid solution. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Adsorption of hydrophobic compounds

Introduce a mixture of solvent containing 50-mM eluent (buffer or aqueous salt solution) and 50 % (v/v) acetonitrile or methanol. Recommended solvent volume to introduce is 5 to 10 times of the column volume.

Keeping the washing solution in the column for a long time will lead to column deterioration. Please replace the washing solution with the eluent immediately after cleaning.

CXpak P-421S

Reverse the column on the HPLC system, flow the eluent at 50 % of ordinary flow rate. 

Cleaning procedure and regeneration of
Chiral column

ORpak CDBS-453

For an efficient cleaning, reverse the flow direction. Set the flow rate at 0.5 mL/min. Introduce 50 mL of acetonitrile or methanol. (Cleaning time: 100 minutes).

Keeping the Washing solution in the column for a long time will lead to column  deterioration. Please replace the Washing solution with the eluent immediately after  cleaning.

For more information, please also read and download the column manuals:

Troubleshooting - Abnormal peak shape

troubleshooting abnormal peak shape

Troubleshooting - Fluctuation in elution time

troubleshooting fluctuation in elution time

Troubleshooting - Fluctuation in baseline

troubleshooting fluctuation in baseline

Troubleshooting - Noise in baseline

troubleshooting noise in baseline

Troubleshooting - Rise in pressure

troubleshooting rise in pressure
bottom of page